Although several methods have been used clinically to assess aortic regurgitation (AR), there is no “gold standard” for regurgitant volume measurement. Magnetic resonance phase velocity mapping (PVM) can be used for noninvasive blood flow measurements. To evaluate the accuracy of PVM in quantifying AR with a single imaging slice in the ascending aorta, in vitro experiments were performed by using a compliant aortic model. Attention was focused on determining the slice location that provided the best results. The most accurate measurements were taken between the aortic valve annulus and the coronary ostia where the measured (Y) and actual (X) flow rate had close agreement (Y = 0.954 × + 0.126, r2 = 0.995, standard deviation of error = 0.139 L/min). Beyond the coronary ostia, coronary flow and aortic compliance negatively affected the accuracy of the measurements. In vivo measurements taken on patients with AR showed the same tendency with the in vitro results. In making decisions regarding patient treatment, diagnostic accuracy is very important. The results from this study suggest that higher accuracy is achieved by placing the slice between the aortic valve and the coronary ostia and that this is the region where attention should be focused for further clinical investigation.
Repository Citation
Chatzimavroudis, George P.; Walker, Peter G.; Oshinski, John N.; Franch, Robert H.; Pettigrew, Roderic I.; and Yoganathan, Ajit P., "Slice Location Dependence of Aortic Regurgitation Measurements with MR Phase Velocity Mapping" (1997). Chemical & Biomedical Engineering Faculty Publications. 99.
https://engagedscholarship.csuohio.edu/encbe_facpub/99
Original Citation
Chatzimavroudis GP, Walker PG, Oshinski JN, Franch RH, Pettigrew RI, Yoganathan AP. Slice location dependence of aortic regurgitation measurements with MR phase velocity mapping. Magnetic Resonance in Medicine. 1997;37:545-551.
Volume
37
Issue
4
DOI
10.1002/mrm.1910370412
Version
Postprint
Publisher's Statement
This is the accepted version of the following article: Chatzimavroudis GP, Walker PG, Oshinski JN, Franch RH, Pettigrew RI, Yoganathan AP. Slice location dependence of aortic regurgitation measurements with MR phase velocity mapping. Magnetic Resonance in Medicine. 1997;37:545-551., which has been published in final form at http://onlinelibrary.wiley.com/doi/10.1002/mrm.1910370412/abstract
Comments
This work was supported by grants from the American Heart Association, Georgia Affiliate, and the Emory/Georgia Tech Biomedical Technology Center.