The Importance of Slice Location on the Accuracy of Aortic Regurgitation Measurements with Magnetic Resonance Phase Velocity Mapping

Document Type

Article

Publication Date

7-1997

Publication Title

Annals of Biomedical Engineering

Abstract

Although several methods have been used clinically to evaluate the severity of aortic regurgitation, there is no purely quantitative approach for aortic regurgitant volume (ARV) measurements. Magnetic resonance phase velocity mapping can be used to quantify the ARV, with a single imaging slice in the ascending aorta, from through-slice velocity measurements. To investigate the accuracy of this technique,in vitro experiments were performed with a compliant model of the ascending aorta. Our goals were to study the effects of slice location on the reliability of the ARV measurements and to determine the location that provides the most accurate results. It was found that when the slice was placed between the aortic valve and the coronary ostia, the measurements were most accurate. Beyond the coronary ostia, aortic compliance and coronary flow negatively affected the accuracy of the measurements, introducing significant errors. This study shows that slice location is important in quantifying the ARV accurately. The higher accuracy achieved with the slice placed between the aortic valve and the coronary ostia suggests that this slice location should be considered and thoroughly examined as the preferred measurement site clinically.

Comments

This work was supported by grants from the American Heart Association, Georgia Affiliate, and the Emory/GeorgiaTech Biomedical Technology Center.

Original Citation

Chatzimavroudis GP, Walker PG, Oshinski JN, Franch RH, Pettigrew RI, Yoganathan AP. The importance of slice location on the accuracy of aortic regurgitation measurements with magnetic resonance phase velocity mapping. Ann Biomed Eng. 1997;25:644-652.

Volume

25

Issue

4

DOI

10.1007/BF02684842

Share

COinS